X

Velp R30800194 Titroline 5000 Automatic Titrator

The TitroLine® 5000 is a very compact titrator for simple routine titrations. Thanks to the standard methods the creating of methods is no longer required. Just load the method and press the start button. GLP compliant results can be documented on a connected printer or USB-memory stick.

Simple automatic titrator for potentiometry

Simple automatic titrator for potentiometry
• Color display that can be easily viewed from a distance and at extreme angles.
• Export the results as PDF or CSV on USB sticks.
• Pre-installed standard methods

• Following titration modes are possible:
• Titrations to pH, mV - endpoint (2 EP)
• Titrations to one equivalence point (pH und mV) with dynamic or linear titration addition
• Manual titrations with the controller
• Dosing tasks
• Number of methods: 5


Typical applications for the TitroLine® 5000 are:
• Salt content in foodstuffs (cheese, soy sauce, ketchup)
• pH- value, total acidity in wine and beverages
• Acid value in bread and bread dough
• Total nitrogen according to Kjeldahl
• Chloride in drinking water and other samples
• pH- value, Alkalinity (p + m-value)
• FTA/TA („FOS/TAC“)
 

TECHNICAL DATA

MEASURING INPUT PH/MV pH/mV-input with elektrode socket according to DIN 19262. BNC via adapter. Reference electrode 1 x 4 mm socket
MEASURING RANGE PH/ACCURACY - 3,0… 17,00 / 0,05 +/- 1 digit
MEASURING RANGE MV/ACCURACY - 1900…1900/ 1,0 +/- 1 digit
MEASURING INPUT TEMPERATURE:  Connector for Pt 1000 / NTC 30 KOhm (Sockets: 2 × 4 mm)
MEASURING RANGE TEMPERATURE - 30 …115 °C / 0,5 K +/- 1 digit
INTERFACES:  2 × USB (1 USB-A and 1 × USB-B), 2 × RS-232-C
DOSING UNITS 20 ml; 50ml also usable 
RESOLUTION (STEPS):  8000 steps
DOSING ACCURACY: according to DIN ISO 8655, part 3:
accuracy: < 0.15 %
precision: < 0.05 %
DISPLAY 3.5 inches - 1/4 VGA TFT graphic colour display with 320 x 240 pixels
HOUSING MATERIAL Polypropylene
DIMENSIONS 135 x 310 x 205 mm (W x H x D), height with dosing unit, without stirrer
FRONT KEYBOARD Polyester
MATERIAL DOSING UNIT/VALVE Valve: PTFE/ETFE;
Cylinder: borosilicate glass 3.3 (DURAN®)
Hoses:  (FEP, blue)
UV Cover: Ultem 
WEIGHT 2 kg
AMBIENT CONDITIONS Ambient temperature: + 10 ... + 40 °C for operation and storage Humidity according to EN 61 010, Part 1: max. relative humidity 80 % for temperatures up to 31 °C, linear decrease down to 50 % relative humidity at a temperature of 40 °C
POWER SUPPLY: 100-240 V; 50/60 Hz, power input: 30 VA
CE-SIGN:  EMC compatibility according to the Council Directive: 2004/108/EG; applied harmonized standards: EN 61326-1:2006
Low-voltage directive according to the Council Directive 2006/95/EG Testing basis EN 61 010, part 1