BINDER FDL 115 Safety Drying Chamber

  • Temperature range: room temperature plus 5 °C to 300 °C
  • APT.line™ preheating chamber technology
  • Controller with time-segment and real-time programming

BINDER FDL 115 Safety Drying Chamber - Internal dimension - W x H x D [mm]: 600 x 435 x 435. Interior Volume [L]: 115

Main characteristics:

  • Temperature range: room temperature plus 5 °C to 300 °C
  • APT.line™ preheating chamber technology
  • Controller with time-segment and real-time programming
  • 2 chrome-plated racks
  • Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
  • Replaceable filter cartridge, Class M6
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Computer interface: RS 422

Benefits:

  • Safety concept meets the DIN EN 1539-2010-08 standard
  • Short drying time with high temperature accuracy and defined ventilation
  • Silicone free and dust-free inner chamber
  • Replaceable fresh air cartridges are easy to change and ensure constant fresh air quality

Important characteristics

  • Temperature range: +10 °C above ambient temperature to +300 °C
  • APT.line™ preheating chamber technology
  • Controller with time-segment and real-time programming
  • 2 chrome-plated racks
  • Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
  • Replaceable filter cartridge, Class M6
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Computer interface: RS 422
  • Standard door gasket made from solvent-resistant FKM for use at up to 200 °C
  • Optional door gasket made from silicone with limited solvent resistance for use at up to 300 °C
Designation

FDL115-230V¹

Option model Standard
Order number 9010-0292
Data
Serie FDL
Performance Data Temperature
Heating-up time to 150 °C [min] 20
Recovery time after 30 seconds door open at 150 °C [min] 12
Temperature fluctuation [± K] 0.8
Temperature range +10 °C above ambient temperature to [°C] 300
Temperature variation at 150 °C [± K] 2.5
Air change data
Air circulation (approx.) [x/min] 20
Volumetric flow rate of exhaust air acc. to EN 1539 at 50 °C [ca. L/min] 400
Solvent quantity
Highest permitted solvent quantity (at T-180 °C, M-100g/mol, U-40g/m3, K=0,5) [g] 6.65
Electrical data
Nominal power [kW] 2.9
Phase (Nominal voltage) 1~
Power frequency [Hz] 50/60
Rated Voltage [V] 230
Housing dimensions not incl. fittings and connections
Depth net [mm] 685
Height net [mm] 805
Width net [mm] 830
Internal Dimensions
Depth [mm] 435
Height [mm] 435
Width [mm] 600
Measures
Interior volume [L] 115
Load per rack [kg] 30
Net weight of the unit (empty) [kg] 90
Permitted load [kg] 60
Wall clearance back [mm] 160
Wall clearance sidewise [mm] 100
Environment-specific data
Energy consumption at 150 °C [Wh/h] 1200
Sound-pressure level [dB(A)] 57
Fixtures
Number of shelves (std./max.) 2/5