Dycometal Climatic Test Chamber CCK-40/48
- Volume: 48 litres
- Temperature range: From -40°C to +150°C
- Humidity range: 10% R.H. to 98% R.H
Dycometal Climatic Test Chamber CCK-40/48
- Robust construction
- Volumes between 81 and 2000 litres
- Temperature range from -70ºC to+180ºC
- Thermal gradients up to 5ºC/min.
- Easy to install, easy to use and reliable
- Compliance with standards: UNE, DIN, MIL, ISO
Features
Construction
- Extremely robust
- Built on a steel tube structure and externally based galvanized steel sheet with oven-dried acrylic paint to RAL-
- Interior Sheet Stainless Steel AISI 304 (18/8), completely sealed and seamless Door painted according to RAL-9010.
- Doors with or without sight
- Thermal insulation panels composed of semi-rigid polyurethane foam of high density and very low thermal conductivity, free of chlorofluorocarbons, asbestos and lead oxide and mercury as stipulated by the Environmental Regulations concerning the installation of thermal-acoustic
- Standard models with capacities from 81 (MINITEST) to 2000
- 4 multidirectional castors in the base of the chamber to facilitate the movement, and/or vibration height-adjustable support (optional depending on model).
- Access port made in PVC with seal cap (different diameters).
- Temperature control by the electronic controller with digital read-actual value and setpoint
- Temperature measuring by PT-100 DIN Class probes
- 2 stainless steel shelves
Control
- Programs numbers stored in the memory:
- 500 segments
- Loops number: from 1 to 999 or infinite
- Optimization of work values by PID parameters in cold and
- Direct humidity in%
- Pt-100 probe
- LED screen with an alphanumerical keyboard that allows visualizing the programmed The screen has an ergonomic position, switch and controls.
- Temperature and humidity monitoring allow programming in section based profiles using ramp and stabilisation.
- The system allows you to save in nonvolatile memory all programmed data, not being affected by power
- Setting alarms. Advise when exceeds a safety limit or for some reason, there is any deviation from the programmed
- Directly reading of the moisture in %HR
- Power Supply Voltage (check depending on the model- phase or three-phase)
- Climatic test Software + PC Cable
Software
Software for monitoring, data logger programming, graphics, alarms and configuration of the instruments.
- Record and storage to
- Display of existing file.
- Creation and modification of the
- Obtaining of Temperature and Humidity records in digital files with the programming of the record
- Record and storage in a file of Essay historical for real values as well as configured
- Essay Graphical Representation, X-Y type, that shows in the screen the setpoint and the real values, as well as the configured
- Essay records available to be converted to ASCII format
TECHNICAL SPECIFICATIONS:
MODELS |
|
CCK- |
40/81 |
40/125 |
40/180 |
40/300 |
40/480 |
40/648 |
40/1000 |
40/1500 |
40/2000 |
|||||||||
|
|
CCK- |
70/81 |
0/125 |
70/180 |
70/300 |
70/480 |
70/648 |
70/1000 |
70/1500 |
70/2000 |
|||||||||
Volume |
|
Liters |
81 |
125 |
180 |
300 |
480 |
648 |
1000 |
1500 |
2000 |
|||||||||
|
Heigth |
mm |
450 |
500 |
600 |
700 |
800 |
900 |
1000 |
1000 |
1000 |
|||||||||
Workspace |
Width |
mm |
450 |
500 |
600 |
700 |
800 |
900 |
1000 |
1000 |
1000 |
|||||||||
|
Depth |
mm |
450 |
500 |
500 |
625 |
750 |
800 |
1000 |
1500 |
2000 |
|||||||||
|
Heigth |
mm |
1235 |
1500 |
1800 |
1950 |
2020 |
2100 |
2100 |
2200 |
2200 |
|||||||||
External dimensions |
Width |
mm |
980 |
1000 |
900 |
1000 |
1100 |
1330 |
1290 |
1290 |
1330 |
|||||||||
|
Depth |
mm |
1050 |
970 |
1100 |
1220 |
1300 |
1450 |
1505 |
2225 |
2650 |
|||||||||
Thermal test |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Minimum temperature |
|
°C |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
Maximum temperature |
|
ºC |
|
|
|
|
|
|
|
|
+180 |
|
|
|
|
|
|
|
|
|
Cooling temperature change rate (1) |
|
°C/min |
2 |
2 |
2 |
2 |
3 |
2 |
3,5 |
2,5 |
3 |
2,5 |
3 |
2,5 |
3 |
2,5 |
2,5 |
2 |
2,5 |
2 |
Heating temperature change rate (1) |
|
°C/min |
3,5 |
3 |
3,5 |
3 |
3,5 |
4 |
4 |
4 |
4 |
4 |
4 |
4 |
4 |
4 |
3,5 |
3,5 |
3,5 |
3,5 |
Temperature stability |
|
ºC |
|
|
|
|
|
|
|
±0,1 ºC to ±0,3ºC |
|
|
|
|
|
|
|
|||
Temperature uniformity |
|
ºC |
|
|
|
|
|
|
|
±0,5 ºC to ±1,5ºC |
|
|
|
|
|
|
|
|||
Maximum compensation (2) |
|
KW |
1,2 |
0,7 |
1,2 |
0,7 |
2 |
1,5 |
2 |
1,5 |
2 |
2 |
2 |
2 |
2,5 |
2 |
3,5 |
3 |
3,5 |
3 |
Climatic test |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Temperature range |
|
ºC |
|
|
|
|
|
|
|
+10ºC to +90ºC |
|
|
|
|
|
|
|
|||
Humidity range |
|
%RH |
|
|
|
|
|
|
|
|
10% to 98% |
|
|
|
|
|
|
|
|
|
Dewpoint - Type 1 |
|
ºC |
|
|
|
|
|
|
|
6,5ºC to 86ºC |
|
|
|
|
|
|
|
|||
Dewpointo - Type 2 |
|
ºC |
|
|
|
|
|
|
hasta -5ºC (limited in time) |
|
|
|
|
|
|
|||||
Humidity stability |
|
%RH |
|
|
|
|
|
|
|
±1%RH to ±3%RH |
|
|
|
|
|
|
|
|||
Maximum compensation (3) |
|
KW |
0,1 |
0,25 |
0,1 |
0,25 |
0,4 |
0,4 |
0,4 |
0,4 |
0,4 |
0,4 |
0,5 |
0,5 |
0,5 |
0,5 |
0,5 |
0,5 |
0,5 |
0,5 |
Power |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Power (kW approx) |
|
kW |
4,3 |
6 |
5,2 |
7,5 |
6,1 |
7,8 |
7,8 |
9 |
9,5 |
10,1 |
10,7 |
11,8 |
12 |
13,4 |
16 |
18 |
20 |
22,9 |
Weight |
|
Kg |
220 |
250 |
250 |
275 |
400 |
425 |
550 |
575 |
700 |
750 |
850 |
920 |
1000 |
1080 |
1120 |
1200 |
1280 |
1340 |
Voltage |
|
V |
|
|
|
|
|
|
|
|
400V 3N~PE |
|
|
|
|
|
|
|
|
|
Condenser |
|
|
|
|
|
|
|
|
|
|
Air cooled |
|
|
|
|
|
|
|
|
|
Noise level (dB aprox.) |
|
dB |
|
|
|
|
|
|
|
|
<60 |
|
|
|
|
|
|
|
|